| TaqMan - Molecular Beacons - Fluorescent Molecular Probes |
| |||||
|
Introduction
Molecular beacons are oligonucleotide probes that can report the presence of specific nucleic acids in homogenous solutions (Tyagi S, Kramer FR. Molecular beacons: probes that fluoresce upon hybridization, Nature Biotechnology 1996; 14: 303-308.) They are useful in situations where it is either not possible or desirable to isolate the probe-target hybrids from an excess of the hybridization probes, such as in real time monitoring of polymerase chain reactions in sealed tubes or in detection of RNAs within living cells. Molecular beacons are hairpin shaped molecules with an internally quenched fluorophore whose fluorescence is restored when they bind to a target nucleic acid (Figure 1). They are designed in such a way that the loop portion of the molecule is a probe sequence complementary to a target nucleic acid molecule. The stem is formed by the annealing of complementary arm sequences on the ends of the probe sequence. A fluorescent moiety is attached to the end of one arm and a quenching moiety is attached to the end of the other arm. The stem keeps these two moieties in close proximity to each other, causing the fluorescence of the fluorophore to be quenched by energy transfer. Since the quencher moiety is a non-fluorescent chromophore and emits the energy that it receives from the fluorophore as heat, the probe is unable to fluoresce. When the probe encounters a target molecule, it forms a hybrid that is longer and more stable than the stem and its rigidity and length preclude the simultaneous existence of the stem hybrid. Thus, the molecular beacon undergoes a spontaneous conformational reorganization that forces the stem apart, and causes the fluorophore and the quencher to move away from each other, leading to the restoration of fluorescence.
Figure 1. Operation of molecular beacons. On their own, these molecules are non-fluorescent, because the stem hybrid keeps the fluorophore close to the quencher. When the probe sequence in the loop hybridizes to its target, forming a rigid double helix, a conformational reorganization occurs that separates the quencher from the fluorophore, restoring fluorescence. In order to detect multiple targets in the same solution, molecular beacons can be made in many different colors utilizing a broad range of fluorophores (Tyagi S, Bratu DP, Kramer FR. Multicolor molecular beacons for allele discrimination, Nature Biotechnology 1998; 16: 49-53.) DABCYL a non-fluorescent chromophore, serves as the universal quencher for any fluorophore in molecular beacons. Owing to their stem, the recognition of targets by molecular beacons is so specific that single-nucleotide differences can be readily detected. Molecular Beacon Example Sequence Fluorophore at 5' end; 5'-GCGAGCTAGGAAACACCAAAGATGATATTTGCTCGC -3'-DABCYL The underlined sequence at the 5' and 3' ends identifies the arm sequences that are complementary. The length of the probe sequence (10-40 nt) is chosen in such a way that the probe target hybrid is stable in the conditions of the assay. The stem sequence (5-7 nt) is chosen to ensure that the two arms hybridize to each other but not to the probe sequence. Folding of the designed sequence with the help of a computer program can indicate whether the intended stem-and-loop conformation will occur. The computer program can also predict the melting temperature of the stem
Signal to background ratio
1. Determine the fluorescence (Fbuffer) of 200 µl of molecular
beacon buffer solution using 491 nm as the excitation wavelength and 515 as the
emission wavelength. If the fluorophore is not fluorescein, chose wavelengths
that are optimal for the fluorophore in the molecular beacon.
Thermal denaturation profiles
1. Prepare two tubes containing 50 µl of 200 nM molecular beacon dissolved in
3.5 mM MgCl2 and 10 mM Tris-HCl, pH 8.0 and add the oligonucleotide
target to one of the tubes at a final concentration of 400 nM.
Real time monitoring of polymerase chain reactions
Procedure
1. Set up six 50 µl reactions so that each contains a different number of
targets, 0.34 µM molecular beacon, 1 µM of each primer, 2.5 units of Amplitaq
Gold DNA polymerase (Perkin Elmer), 0.25 mM of each deoxyribonucleotide, 3.5 mM
MgCl2, 50 mM KCl, and 10 mM Tris-HCl, pH 8.0.
Troubleshooting
Low signal-to-background ratio.
The assay medium may contain insufficient salt. There
should be at least 1 mM MgCl2 in the solution, in order to ensure
that the stem hybrid forms. The molecular beacon may fold into an alternate conformation that results in a
sub-population that is not quenched well. Change the stem sequence (and probe
sequence, if necessary) to eliminate that possibility.
Incomplete restoration of fluorescence at low temperatures.
If the stem of a molecular beacon is too strong, at low temperatures it may remain closed while the probe is bound
to the target. This may happen inadvertently if the probe sequence can
participate in the formation of a hairpin that results in a stem longer and
stronger than originally designed. Change the sequence at the edges of the probe
and the stem sequence to avoid this problem.